Test ID NBLD0716 Synthetic Glucocorticoid Screen, Random, Urine
Useful For
Confirming the presence of the listed synthetic glucocorticoids
Confirming the cause of secondary adrenal insufficiency
Specimen Type
UrineOrdering Guidance
Specimen Required
Supplies: Urine tubes, 10 mL (T068)
Container/Tube: Plastic, 10-mL urine tube
Specimen Volume: 5 mL
Collection Instructions:
1. Collect a random urine specimen.
2. No preservative.
Specimen Minimum Volume
0.6 mL
Specimen Stability Information
Specimen Type | Temperature | Time |
---|---|---|
Urine | Frozen | 14 days |
Reject Due To
All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.Day(s) Performed
Wednesday
Report Available
5 to 11 daysMethod Name
Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)
Reporting Name
Synthetic Glucocorticoid Screen, UPerforming Laboratory

CPT Code Information
80299
Reference Values
Negative
Cutoff concentrations
Betamethasone: 0.10 mcg/dL
Budesonide: 0.20 mcg/dL
Dexamethasone: 0.10 mcg/dL
Fludrocortisone: 0.10 mcg/dL
Megestrol acetate: 0.10 mcg/dL
Methylprednisolone: 0.10 mcg/dL
Prednisolone: 0.10 mcg/dL
Prednisone: 0.10 mcg/dL
Triamcinolone acetonide: 0.10 mcg/dL
Values for normal patients not taking these synthetic glucocorticoids should be less than the cutoff concentration (detection limit).